FLUKE 707

SKU: FLUKE 707

BRAND: Fluke


Current Loop Calibrator 24MA

Overview

Features

Specification

Current Measurement:
    • Range: 0 to 24 mA
    • Resolution: 0.001 mA
    • Accuracy: 0.015% Rdg + 2 LSD


Current Sourcing:
    • Range: 0 to 20 mA or 4 to 20 mA
    • Accuracy: 0.015% Rdg + 2 LSD
    • Drive Capability: 1200 O; 950 O at 20 mA in HART mode


Current Simulation:
    • Range: 0 to 20 mA or 4 to 20 mA
    • Accuracy: 0.015% Rdg + 2 LSD
    • Ext. loop voltage requirement: 12-30 V


Loop Power:
    • While measuring mA, 24 V


Voltage measurement:
  • Range: 0 to 28 V DC
  • Resolution: 1 mV
  • Accuracy: 0.015% Rdg + 2 LSD


Environmental Specifications
  • Operating Temperature: -10ºC to 55ºC
  • Storage Temperature: -30ºC to 60ºC
  • Humidity (Without Condensation): 0 to 90% (0ºC to 35ºC); 0 to 70% (35ºC to 55ºC)
  • Temperature Coefficient: -10ºC to 18ºC, 28ºC to 55ºC, ±.005% of range per ºC


Safety Specifications
  • Safety: CSA C22.2 No. 1010.1:1992
  • EMC: EN50082-1:1992 and EN55022:1994 Class B
  • Maximum voltage: 30V


Mechanical & General Specifications
  • Size: 152 x 70 x 41 mm
  • Weight: 224 g (340 g including holster)
  • Batteries: One 9V battery Alkaline
  • Warranty: Three years
  • Shock & Vibration: per MIL-T-28800 for a Class 2 Instrument
  • Display current and % of span: Yes
  • Auto step, auto ramp: Yes
  • Span Check: Yes